ULTHERAPY PRIME

Guía de códigos Ultherapy

Códigos Ultherapy comunes, causas posibles y primeras revisiones para documentar fallos y solicitar reparación.

Enviar consulta
Equipo Ultherapy Prime
01

Detener primero

Ante temperatura, hardware, energía o botón STOP, no fuerce escaneos repetidos.

02

Registrar código

Fotografíe la pantalla y registre código, modelo de transductor, líneas restantes y pasos.

03

Revisar accesorios

Pieza de mano, transductor, USB, ventilación y contacto son puntos iniciales comunes.

CÓDIGOS DE MENSAJE

Códigos Ultherapy comunes

Referencia inicial. No sustituye IFU, manual ni revisión in situ. Si el error se repite, detenga el uso.

Code Área Posible significado Primera revisión
B Pieza de mano Internal handpiece temperature is too high. Pause use and let the handpiece cool down; check for long continuous operation or poor ventilation.
C Unidad de control The control unit detected a hardware event and stopped. Restart as instructed on screen; if it repeats, record the code and arrange inspection.
E Comunicación Communication halted during initialization. Restart the system and confirm the boot process completes; avoid power loss or removing accessories mid-process.
G Unidad de control The control unit detected a hardware event and stopped. Restart and observe whether it recurs; keep context and photos if it does.
H Transductor Transducer motion was not detected. Stop scanning first, then remove and reinsert the transducer; confirm it is latched into the handpiece.
I Comunicación The control unit detected a communication event. Restart and observe; if accompanied by handpiece or transducer issues, record accessory status.
J Pieza de mano Handpiece communication halted. Restart and check whether the handpiece cable and port are secure.
K Software Software halted due to an event. Restart the system; if it persists, avoid repeated attempts and arrange inspection.
L Transductor The transducer has no remaining lines. Replace the transducer before continuing; record the model and remaining-line status.
M Pieza de mano Handpiece motion halted. Check that the handpiece and transducer are correctly installed and locked; reinstall and test again.
N USB USB drive connection or read/write issue. Confirm USB format, free space, and insertion status; do not remove it during system read/write.
P Unidad de control The control unit detected a hardware event and stopped. Restart; if it persists, record the code and arrange hardware inspection.
S Parada de emergencia The red STOP button was pressed. Confirm STOP button status, rule out accidental press, then restart as instructed.
T Transductor Internal transducer temperature is too high. Pause use and let the transducer cool; use another transducer if needed.
U Unidad de control Internal control unit temperature is too high. Check ambient temperature and ventilation; let the main unit cool before testing again.
V Entrega de energía Energy delivery halted, possibly due to excessive reflected power. Resume scanning as instructed; if it repeats, try another transducer and arrange inspection.
W Transductor The transducer is unauthorized or not accepted by the system. Replace with a transducer suitable for the device; do not force continued operation.
X Transductor The system cannot read the transducer. Remove and reinsert the transducer; check that the contact area is clean and dry.
Ultherapy Prime 探頭

TRANSDUCER CHECK

Para errores de transductor, revise primero contacto y bloqueo

H, L, M, T, V, W y X suelen relacionarse con transductor, pieza de mano o energía. Fotografíe contacto, bloqueo, líneas y código.

THERMAGE FLX MANUAL NOTES

Notas rápidas Thermage FLX

  • When a Thermage FLX system error appears, record the code and screen message first, then follow on-screen instructions.
  • If the activation button does not respond, confirm the handpiece is connected and the footswitch has not taken over activation control.
  • If the touchscreen does not accept touches, press the center of the button and check whether the screen is locked.
  • If the return pad heats up, stop operation and check pad placement and cable connection before continuing.
  • If the handpiece overheats abnormally or the device has visible physical damage, stop use and arrange professional inspection.

REFERENCE IMAGES

Imágenes originales de códigos

Imágenes conservadas para comprobación manual; use la tabla HTML superior para lectura normal.

Ultherapy 錯誤代碼 B 到 I Ultherapy 錯誤代碼 J 到 T Ultherapy 錯誤代碼 U 到 X

El resumen se basa en los documentos aportados y solo sirve para revisiones iniciales y comunicación de reparación.